EasyManua.ls Logo

Rohde & Schwarz ZNB20 - Calculate<Ch>:Transform:vnetworks:balanced:deembedding<Logpt; Calculate<Ch>:Transform:vnetworks:balanced:deembedding<Logpt>: Parameters:l<Cmp; Calculate<Ch>:Transform:vnetworks:balanced:deembedding<Logpt>:Tndefinition

Rohde & Schwarz ZNB20
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
952User Manual 1173.9163.02 ─ 54
SINCreasing
Swapped increasing port sequence (high port numbers towards
VNA, low port numbers towards DUT)
<arg1> <block_data>
Content of a Touchstone file (*.s2p or *.s4p) in IEEE488.2
Block Data Format.
Usage: Setting only
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:G<Cmp> <CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:G<Cmp>? <CircuitModel>
Specifies the conductance value G<Cmp> in the different circuit models for balanced
port deembedding.
In the query form, the <Conductance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<LogPt> Logical port number (balanced port)
<Cmp> Number of conductance in circuit model. The total number of
conductances depends on the selected circuit model.
Parameters:
<Conductance> Conductance G<Cmp> for the specified circuit model.
Range: -1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST: 0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel> STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Possible circuit models (character data); see Table 8-6.
Manual operation: See "Network" on page 607
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:L<Cmp> <CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:L<Cmp>? <CircuitModel>
Specifies the inductance value L<Cmp> in the different circuit models for balanced port
deembedding.
In the query form, the <Inductance> parameter must be omitted. The command returns
the inductance value for the specified circuit model.