MMEMory:LOAD:CORRection:MERGe <Channel>[, <CalGroupFile>,
<CalGroupFile>...]
Merges (activates) several cal group files for channel no. <Channel> so that the query
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine? returns a list of all
merged calibration types (equivalent to the calibration pool list in the "Calibration Man-
ager" dialog). The merged cal group files can be stored to a common file (see exam-
ple).
Note that the calibrations to be merged must be based on the same frequency grid
(identical frequency sweep points).
Setting parameters:
<Channel> Channel number of an existing channel
<CalGroupFile>
<CalGroupFile> String parameters with the names of the merged cal group files.
Cal group files must have the extension *.cal. The file exten-
sions must be specified as part of the string parameters. In con-
trast the directory path must not be specified; the analyzer
always uses the default path
C:\Users\Public\Documents\Rohde-Schwarz\Vna\
CALibration\DATA.
Example:
SENS1:CORR:COLL:METH:DEF 'Test', FRTRans, 1, 3
Select a bidirectional transmission normalization between ports
1 and 3 as a calibration type for channel 1.
CORR:COLL:SAVE:SEL:DEF; :MMEM:STOR:CORR 1,
'P1-P3.cal'
Create a default calibration data set for the selected calibration
type and store the data to a cal group file.
SENS1:CORR:COLL:METH:DEF 'Test', FRTRans, 1, 4
Select a bidirectional transmission normalization between ports
1 and 4 as a calibration type for channel 1.
CORR:COLL:SAVE:SEL:DEF; :MMEM:STOR:CORR 1,
'P1-P4.cal'
Create a default calibration data set for the selected calibration
type and store the data to a cal group file.
CORR:COLL:METH:DEF?
Query the active calibrations for channel 1. The response is
FRTR0104 (the last data set stored).
MMEM:LOAD:CORR:
MERGE 1, 'P1-P3.cal', 'P1-P4.cal'; :CORR:COLL:
METH:DEF?
Merge the two calibration types and query the active calibrations
again. The response is FRTR0103,FRTR0104.
MMEM:STOR:CORR 1, 'Merged.cal'
Store both sets of calibration data to a common cal group file.
Usage:
Setting only
SCPI Command Reference