Command Reference
R&S
®
ZNB/ZNBT
1091User Manual 1173.9163.02 ─ 54
Example:
*RST; :CALC:TRAN:VNET:PPA:EMB:DEF 1,2,3,4
Define a port pair configuration with port pairs (1,2) and (3,4).
CALC:TRAN:VNET:PPA:EMB2:TND STSL
Select the Serial Touchstone .s2p data, shunt L circuit mode for
the second port pair.
MMEM:LOAD:VNET:PPA:EMB2 'C:
\Rohde-Schwarz\Vna\Traces\Test.s2p', PMA
Load a Touchstone file and assign it to the second port pair.
Manual operation: See "File Name <i>/Inc. Seq. <i>" on page 576
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
<TouchstoneFile>[, <Interchange>]
Loads data from a specified two-port (*.s2p) Touchstone file defining a single ended
circuit model for deembedding.
Use
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition to select the adequate circuit model before executing this com-
mand.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:DATA to load circuit data from the remote client.
The query returns the name of the loaded file.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<TouchstoneFile> String parameter to specify the name and directory of the loaded
Touchstone file.
If no path is specified the analyzer searches the current direc-
tory, which can be set and queried using MMEMory:
CDIRectory
<Interchange> FPORts | IPORts | SGATes
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
Example:
CALC:TRAN:VNET:SEND:DEEM:TND FIMP
Select the "Serial .s2p" data circuit model.
MMEM:LOAD:VNET:SEND:DEEM2 'C:\Users\Public
\Documents\Rohde-Schwarz\Vna\VNET\Test.s2p'
Load a Touchstone file and assign it to the physical port no. 2.
Manual operation: See "File Name 1 / Swap Gates" on page 574