EasyManua.ls Logo

Rohde & Schwarz ZNB20 - Page 1190

Rohde & Schwarz ZNB20
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1190User Manual 1173.9163.02 ─ 54
Suffix:
<Ch>
.
Channel number
<DCInp> Number of DC INPUT connector on the rear panel, 1 to 4
Parameters:
<Voltage> DC voltage range. The entered numeric value is rounded up to
±300 mV, ±3 V, ±20 V. Values above +20 V are rounded down.
Default unit: V
Example:
SENSe1:DC2:RANGe 1; :DC2:RANGE?
Select a DC measurement at DC INPUT2 with an expected DC
voltage range of ±1 V. The value is rounded up; the response is
3.
Manual operation: See "Ranges" on page 288
8.3.14.11 [SENSe:]FREQuency...
The [SENSe:]... commands set frequency-related parameters, especially the mea-
surement and display ranges for the different sweep types. The frequency ranges for
the different instrument models are listed below; for more details refer to the data
sheet.
Table 8-16: Frequency ranges of R&S
ZNB/ZNBT analyzers
Frequency settings Start, Stop Center Span
R&S ZNB4 9 kHz to 4.5 GHz >9 kHz to <4.5 GHz 1 Hz to 4.499991 GHz
R&S ZNB8 /
R&S ZNBT8
9 kHz to 8.5 GHz >9 kHz to <8.5 GHz 1 Hz to 8.499991 GHz
R&S ZNB20 /
R&S ZNBT20
100 kHz to 20 GHz >100 kHz to <20 GHz 1 Hz to 19.9999 GHz
R&S ZNBT26 100 kHz to 26.5 GHz >100 kHz to <26.5 GHz 1 Hz to 26.4999 GHz
R&S ZNB40 variant 72 10 MHz to 40 GHz >10 MHz to <40 GHz 1 Hz to 39.99 GHz
R&S ZNB40 variants 82,
83 and 84
R&S ZNBT40
100 kHz to 40 GHz >100 kHz to <40 GHz 1 Hz to 39.9999 GHz
Option R&S ZNB-B1, Bias Tees
If option R&S ZNB-B1 is installed on a R&S ZNB4 or R&S ZNB8, the frequency range
starts at 100 kHz.
Bias Tees are not available for the R&S ZNBT.
[SENSe<Ch>:]FREQuency[:CW]....................................................................................1191
[SENSe<Ch>:]FREQuency:FIXed.................................................................................. 1191
[SENSe<Ch>:]FREQuency:CENTer............................................................................... 1192
[SENSe<Ch>:]FREQuency:CONVersion.........................................................................1192
[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary.......................................................... 1193
SCPI Command Reference