Command Reference
R&S
®
ZNB/ZNBT
1313User Manual 1173.9163.02 ─ 54
<Driver> Generator type (string parameter). The generator type is identi-
cal with the name of the generator driver file (*.gen) stored in
the resources\extdev subdirectory of the analyzer's program
directory. Type the driver file name as shown in the "Add Exter-
nal Generator" dialog, i.e. without the file extension ".gen"
(example: use 'sme02', if the corresponding driver file name is
sme02.gen).
Alternative: The '<driver>' string may also contain the gener-
ator driver file name with its complete path, e.g. 'C:\Program
Files\Rohde-Schwarz\Vector Network
Analyzer\ZNB\resources\extdev\sme02.gen'.
<Interface>
Interface type (string parameter): 'GPIB0', 'VXI-11',
'socket','usb-visa', 'other' ...
<Address> Interface address (string parameter), depending on the interface
type. See Chapter 6.17.4.4, "External Generators Dialog",
on page 725.
<FastSweep> ON | OFF
Optional Boolean parameter, enables or disables the fast sweep
mode.
<ExtReference> ON | OFF
Optional Boolean parameter, sets the analyzer to internal (OFF)
or external (ON) reference frequency.
Example:
SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02',
'gpib0', '21'
Configure an R&S SME02 generator as external generator no.
1, assigning the name "Ext. Gen. 1" and a GPIB address 21.
SYST:COMM:RDEV:GEN1:DEF?
Query the generator configuration. The response 'Ext. Gen.
1','SME02','GPIB0','21',OFF, OFF indicates that the
fast sweep mode is disabled and that the analyzer is set to inter-
nal reference frequency.
SYST:COMM:RDEV:GEN1:DEL
Clear the generator configuration table.
SYST:COMM:RDEV:GEN1:DEF?
Query the generator configuration. The analyzer returns an error
message because the generator no. 1 is no longer configured.
Manual operation:
See "Configured Devices" on page 727
SYSTem:COMMunicate:RDEVice:GENerator:DELete
Clears the configuration table for external generators.
Example:
See SYSTem:COMMunicate:RDEVice:GENerator<Gen>:
DEFine
Usage: Event
SCPI Command Reference