EasyManua.ls Logo

Rohde & Schwarz ZNB20 - Page 974

Rohde & Schwarz ZNB20
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
974User Manual 1173.9163.02 ─ 54
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:L<1|2|3> <CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:L<1|2|3>? <CircuitModel>
Specifies the inductance value L<i> in the different lumped circuit models for port pair
deembedding.
In the query form, the <Inductance> parameter must be omitted. The command returns
the inductance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port pair (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine)
<1|2|3> 1|2|3
Index i of the inductance L<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.
Parameters:
<Inductance> Inductance L<i>
Range: -1 H to 1 H.
Increment: 1 pH (1E-12 H)
*RST: 1 nH (1E-9 H)
Default unit: H
Parameters for setting and query:
<CircuitModel> STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS | SLLS |
GSSL | LSSG | SLGS | SGLS
Circuit model whose inductance L<i> shall be set, see Table 8-6
Example: See CALCulate<Ch>:TRANsform:VNETworks:PPAir:
DEEMbedding<ListId>[:STATe]
Manual operation: See "Network" on page 604
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:R<1|2|3> <CircuitModel>[, <Resistance>]
Specifies the resistance value R<i> in the different lumped circuit models for port pair
deembedding.
In the query form, the <Resistance> parameter must be omitted. The command returns
the resistance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port pair (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine)
SCPI Command Reference

Table of Contents

Related product manuals