2-6
2.3.3 Microscope
Insert the microscope mount arm into the end of the microscope rod that extends from
the front panel, and clamp firmly in place with the provided set screw. Install the
microscope into the mount arm. The microscope can be used at any angle between 60
degrees and 36 degrees to the horizontal.
2.3.4 Sample Holder
Place the sample holder (or manipulator/test piece holder combination) on the base
plate. After inserting a sample, slide the sample holder into testing position by hand. The
weight of the sample holder is enough to secure it to the base plate during testing.
2.3.5 Pull Hooks
Carefully, open the pull hook shipping tube by twisting the cap while pulling the cap and
tube apart gently. Pull hooks are fragile and can easily be damaged by contact, handle
the hook by the aluminum body ONLY. Loosen the set screw and remove the hook from
the cap. Carefully slide the hook onto the pull rod and tighten the set screw.
2.3.6 Mouse
Attach the mouse cable to the connector labeled “MOUSE” on the back panel. Place the
mouse on a mouse pad. Keep the mouse pad clean to ensure proper functioning of the
mouse.
2.3.7 Options
Attach the optional keypad to the connector labeled “KEYPAD” on the back panel. The
optional printer is attached to the connector labeled “PRINTER” on the back panel.
2.3.8 Initial Calibration
Acclimatize the system to the installation environment by checking the system calibration
and calibrating the system if necessary. Perform this procedure whenever the system is
moved or the environmental conditions surrounding the system change.