EasyManua.ls Logo

Royce 610 - Page 27

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-4
TEST STOPS AT
This selection allows the operator to set the percentage of maximum test force
below which the test is concluded. The default setting is edited by pressing the left
or right mouse buttons to increase or decrease the value, then press the center
mouse button to accept the value.
HOOK Z SPEED
This selection allows the operator to set the Z axis speed of the test hook as
controlled by the left and right mouse buttons. The speed can be adjusted from a
minimum of 1 (50 mils per second) to a maximum of 10 (500 mils per second) with
a default value of 5 (250 mils per second). The default setting is edited by
pressing the left or right mouse buttons to increase or decrease the value, then
press the center mouse button to accept the value.
HOOK θ
θθ
θ SPEED
This selection allows the operator to set the rotational speed of the test hook as
controlled by positional movement of the mouse. The default setting is edited by
pressing the left or right mouse buttons to increase or decrease the value, then
press the center mouse button to accept the value.
CODE 1:
This selection allows the operator use the keypad to edit the description of fail
code #1.
CODE 2:
This selection allows the operator use the keypad to edit the description of fail
code #2.
CODE 3:
This selection allows the operator use the keypad to edit the description of fail
code #3.
CODE 4:
This selection allows the operator use the keypad to edit the description of fail
code #4.
CODE 5:
This selection allows the operator use the keypad to edit the description of fail
code #5.
CODE 6:
This selection allows the operator use the keypad to edit the description of fail
code #6.