3-3
GROUP NUMBER
This selection allows the operator to enter a group identification number for the
current sample group. Use the keypad to enter a group number designation.
LOWER HT
This selection allows the operator to set the lower limit for test hook travel. To set
the lower height, use the left or right mouse buttons to position the hook at the
desired lower height position, then press the center button to accept the value.
SEARCH HT
This selection allows the operator to set the test hook search height. To set the
search height, use the mouse buttons to position the hook at the desired search
height position, then press the center mouse button to accept the value.
RETURN TO
This selection allows the operator to choose the height to which the hook returns
upon completion of a test. The choices are “START HT” or “SEARCH HT”.
ALARM
This selection allows the operator to set an alarm based on the time of day. The
alarm time is displayed as a 24 hour value and can be set to any time from 00:00
to 23:59. An audible “double beep” sounds at the alarm time.
SPEED
This selection allows the operator to set the Z axis speed of the test hook as
controlled by the system during testing. The speed can be adjusted from a
minimum of 4 mils per second to a maximum of 240 mils per second with a default
value of 100 mils per second. The default setting is edited by pressing the left or
right mouse buttons to increase or decrease the value, then press the center
mouse button to accept the value.
NDT HOLD
This selection allows the operator to enter a hold time for the non-destruct test
pull. The default setting is edited by pressing the left or right mouse buttons to
increase or decrease the value, then press the center mouse button to accept the
value.
SEMI AUTO
This selection allows the operator to toggle the semi automatic testing mode on or
off. Press the center mouse button to toggle the setting.
MUST FAIL CODE
This selection allows the operator to set the system to prompt for a fail code and
wait until the code is entered before resuming test mode. Press the center mouse
button to toggle the setting.