EasyManua.ls Logo

Sanyo MDF-U500VX - Parameter Settings and Defaults; Parameter Setting Ranges

Sanyo MDF-U500VX
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 37 -
15. Setting parameter and factory defaults
Chamber temp.
-90㷄䌾-50
(Default: -80 )
High temp. alarm
+5㷄䌾+20
(Default: +10 )
Low temp. alarm
-5㷄䌾-20
(Default: -10 )
Alarm delay time
0min.15min.
(Default: 15min.)
Ring Back
1min.~99min. (0:OFF)
(Default: 30min.)
Door delay time
1min.~ 15min.
(Default: 2min.)
Compressor differential value
ON: Set temp.+0.4
OFF: Set temp.-0.4
Auto return
Returns to TopScreen if there are no key operations for 90
seconds.
Keylock password Default: 0000
Cap.tube heater intervals
Once per 24 hours
(Heater is not energized)
Energizing period of cap. tube heater
6min.~15min.
(Default: 8min.)
Log intervals
2min.~30min.
(Default: 15min.)
Compressor delay time 3 min.
Compressor delay time when power
returns from power failure
2min.~15min.
(Default: 3min.)
When condenser temp is higher than +60, Compressor
stops running.
Compressor protection
When comp sensor temp is AT+10, Compressor starts
running
Chamber temp. initial offset
0.0, changeable
Auto MENU OFF
MENU button disappears if there are no button operations
for 60sec.

Related product manuals