31
F08 To adjust 0 of temperature sensor setting range: -9.9℃~+9.9℃)
In F17: It is functioned in model code 002(MDF-U537), 003(MDF-U537D).
Return to PV display in model code 001(MDF-U333)
Setting method: In 08” displayed, press SET key to display00.0”initial value.
Set temperature with key and key and press SET key
again to memorize set value. Input is done by differential value.
Ex When actual temperature of 1/2h chamber is shown –28.5and digital
display is shown -30, add +1.5 to value in F08.
Calculation = (-28.5-(-30)) = +1.5
key enables to change only “0” and “-“ in second digit and to change
“09 in first digit and first decimal
place.
key enables to shift the second digit the first digit
the first decimal placethe second digit.
12 Temperature of temperature sensor is displayed as per 0.1.
From second digit to first decimal point is shown in display.
It is not shown “- in display when temperature is lower than -20.
(Ex. Actual temperature -35 Display shows 35.0)
When temperature is within the range of -19.9℃~-0.1,- is displayed.
Range : -72.0+83.0℃(±9.9℃)
13 Temperature of compressor protect sensor is displayed as per 1.
Range: -72+163
14 Temperature sensor (solenoid valve control sensor for MDF-U537, lower compartment
temperature control sensor for MDF-U537D)
Temperature of each sensor is displayed as per 0.1
From second digit to first decimal point is shown in display.
It is not shown “- in display when temperature is lower than -20.
(Ex. Actual temperature -35 Display shows 35.0)
When temperature is within the range of -19.9℃~-0.1, ”-“ is displayed.
Range : -72.0+83.0