- 14 -
F21: Input service code “384” in F06 before utilizing F21.
Communication ID can be set in F21.
Setting range; 000 to 255
Press CALL key to memorize the value and chamber temperature
display will appear.
If there are no key operations for approx. 40seconds, chamber
temperature display will appear.
F22: Input service code “384” in F06 before utilizing F22.
Communication mode can be set in F22.
Setting range; “0” for 1
st
digit
“0” or “1” or “2” for 2
nd
digit
“0” or “1” for 3
rd
digit
Factory default setting is “000”.
Press CALL key to memorize the value and chamber temperature
display will appear.
If there are no key operations for approx. 40seconds, chamber
temperature display will appear.
7. Timer mode
If timer mode operation is complete, a unit will keep operation
displaying --:-- in same setting temperature.