- 38 -
15
STABLE RUNNING WITH DELAY TIME
Table 2 shows one of applications. Perform key
operations in the sequence indicated in the table.
The example in the table is the delay time that is 30
minutes, the chamber temperature is 50
o
C, the
running time is 1 hour and the timer mode is Timer
Mode C.
Table 2: Basic operation sequence (Example: Chamber temperature; 50
o
C,
Delay time; 30 minuets, Running time; 1 hour, Timer mode; C)
Description of operation Key operated Indication after operation
1 Press the power key.
The current chamber temperature is displayed.
2 Press the delay timer key.
The current delay time is displayed.
3
Press the delay timer key to reset
the delay time.
Reset the delay time.
4
Set the delay time to 00:30 with the
digit shift key and the numeric
value shift key.
5 Press the delay timer key.
The current chamber temperature is displayed.
6 Press the call key.
The fourth digit of the indicator flashes.
7
Set the temp. to 50.0 and running
time to 01:00 with the digit shift key
and numeric value shift key.
8
Press the timer mode key to select
the time mode C.
By pressing the key, the timer mode C lamp is
lighted.
9 Press the run/stop key.
The setting mode is finished and the unit runs.
Note:
If there is no need to change the set value in each setting mode, press the call key (CALL) to activate the
next mode.
If the unit runs every time, set the running time --:-- by pressing the numerical value shift key at the fourth
digit of the time indicator.
Press the Run/Stop key (RUN/STOP) when cancellation is necessary while running.
If you don’t need the delay time, set the delay time 00:00.
If you need to change the set value, key operate after the unit stops.
When no key is pressed within 45 seconds, displayed number is memorized, setting is over and indicator
displays the current value.
POWER
DELAY
CALL
RUN
STOP
DELAY
DELAY
TIMER
MODE
30 min.
50
o
C
1 hour
Buzzer