Note) Following data are the reference only, so they do not assure product’s performance.
1. Pull-up test
z Test condition: AT20Υ with 1 Plastic dish (ȭ90mm) placed in the center of chamber
MIR-162
SV
When unit is stabilized (Υ) Overshooting (Υ)
Time to reach final
value +/-1Υ (Min.)
Display 1/2H air Dish Display 1/2H air Dish Display 1/2H air
25Υ
25.0 24.8 24.9 2.2 1.2 0.7 43 38
30Υ
30.0 29.8 29.9 1.3 0 0 29 22
37Υ
37.0 36.5 36.5 0.6 0 0 21 52
50Υ
50.0 49.3 49.4 0.4 0 0 39 72
60Υ
60.0 59.2 59.3 0.4 0 0 43 85
MIR-262
SV
When unit is stabilized (Υ) Overshooting (Υ)
Time to reach final
value +/-1Υ (Min.)
Display 1/2H air Dish Display 1/2H air Dish Display 1/2H air
25Υ
25.0 25.1 25.0 1.3 0.2 0 35 22
30Υ
30.0 30.2 30.0 0.4 0 0 19 58
37Υ
37.0 37.3 37.3 0.4 0 0 24 82
50Υ
50.0 50.3 50.3 0.5 0 0 49 97
60Υ
60.0 60.6 60.6 0.5 0.1 0.1 68 104
Test data