14 Error Display and Remedy
For Error Display and Remedy please refer Table 4
CODE ERROR DISPLAY ERROR DESCRIPTION TROUBLESHOOTING
1 PrOP Probes are OPEN circuited Check the probes with
multimeter
2 PrSC Probes are SHORT cir-
cuited
Check the probes with
multimeter
3 ECAL Calibration Error Calibration is wrong,
Please recalibrate correctly
4 PrHI Over Capacitance Tank and probe dimen-
sions are not matched
5 PrLO Under Capacitance Tank and probe dimen-
sions are not matched
6 RFOP Internal reference got open
circuited
Internal fault in sensor
7 RFSC Internal reference got short
circuited
Internal fault in sensor
8 RFHI Internal reference satu-
rated
Internal fault in sensor
9 RFLO Internal reference negligi-
ble
Internal fault in sensor
10 OSC Oscillations stopped Internal fault in sensor
Table 4: Error Display and Remedy