| 5
New Features Compared to the BP:  New, different housings (see Overview, Page 6) 
Procedure:  Checking the balance/scale  1.  Check the repeatability 
  Balance/Scale with strain gauge system (standard deviation, see page 18)
  Balance/Scale with double lever system (standard deviation, see page 23)
  Balance/Scale with monolithic weigh cell (standard deviation, see page 28)
  Balance/Scale with forked lever system (standard deviation, see page 35)
2.  Checking the off-center loading error and adjust as required
  Balance/Scale with strain gauge system (see page 18)
  Balance/Scale with double lever system (see page 24)
  Balance/Scale with monolithic weigh cell (see page 29)
  Balance/Scale with forked lever system (see page 36-37)
3.  Adjusting the sensitivity with internal calibration weight
  Balance/Scale with monolithic weigh cell (see page 32)
4.  Adjusting the sensitivity with external calibration weight 
  Balance/Scale with strain gauge system (see page 21)
  Balance/Scale with double lever system (see page 26)
  Balance/Scale with monolithic weigh cell (see page 31)
  Determine the internal calibration weight as required (see page 31)
  balance/scale with forked lever system (see page 38) 
5.  Adjusting the linearity is preferably performed using the sequential method, or 
with calibrated weights,
  Balance/Scale with strain gauge system (see page 22)
  Balance/Scale with double lever system (see page 27)
  Balance/Scale with monolithic weigh cell (see page 33)
  Check the internal linearization weights as required (see page 34)
  Balance/Scale with forked lever system (see page 39)
 
Activating the BPI Mode/  Perform only before working with the SARTOCAS Program or PSION; for calibration/
Deactivating write-protect  adjustment. Linearization and programming a data record when exchanging the 
main PCB.
Reset write-protect  After working with the SARTOCAS program or PSION, always make sure to conclude 
with the „Close“ function (returning the balance from BPI back to SBI mode), 
otherwise write-protect is not set, and the balance/scale remains in the BPI mode 
(see Page 11). If all else fails, reset the SBI mode by pressing menu reset (9-1°).
Exchanging the Main PCB  When ordering replacement PCBs, always state the model and the serial number, that 
is the only way that a new preprogrammed PCB set can be delivered.
Opening the housing  -  Unscrew the 1 (sealed) screw on the reverse side and slide the upper half of 
the housing back.
  -  After reclosing the balance/scale, the service technician or service organization 
must reaffi x another control seal!
Triangular weighing pan  -  Check the off-center loading error according to the OIML Recommendation R76, 
see 4-point test (Pages 18; 24; 29)
  -  when carrying out adjustment/calibration work using the 3-point test (page 30)
Blue backlighting   -  Is adjustable and can be switched off
Service software  -  TradeCAS or SartoCAS starting with Version 1.44
    Psion CAS Version 4.9 and later.
    Do not use older software versions!
Adjusting the off-center load  -  On strain gauge, Angled lever and double lever systems perform by fi ling on 
monolithic weigh cell with adjustment screws
Service Guide