32 |
 -  Make sure that the balance/scale operating menu is set to the code „1 9 3 - 
internal calibration/adjustment accessible „ (see Page 13-14 „ Operating Menu 
Settings „).
-  Slide the access lock switch to the left. It is accessible via an opening on the 
back of the balance/scale.
-  If you have done so already, perform external calibration/adjustment (see Page 
31).
-  Turn off the balance/scale by pressing e.
-  Turn the balance/scale back on by pressing e and while simultaneously 
holding down the q key until the readout shown on the left appears.
- Press the t key to tare the balance/scale; then press q to activate the 
„overwrite internal calibration weight“ function.
-  The built-in, motorized internal calibration weight is applied, then removed 
automatically; the weight value is overwritten; and the balance/scale returns to 
the normal weighing mode.
- Press the q key to check the external span adjustment once again.
Internal  Span Adjustment
-  Make sure that the menu code „1 9 3 - internal calibration/adjustment acces-
sible“ is set (see Page 13 -14 Operating Menu Settings).
- Press the t key to tare the balance/scale; then press q to activate the 
calibration/adjustment function..
-  The built-in, motorized internal calibration weight is applied, then removed 
automatically; the weight value is overwritten; and the balance/scale returns to 
the normal weighing mode.
-  Slide the access lock switch back to the right. Replace the capped plug to close 
the opening. 