Sartorius CPA 12001S - Internal Calibration Procedure Steps; Step-by-Step Internal Calibration

Sartorius CPA 12001S
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
Internal Calibration
Step Key (or instruction) Display
1. Zero the balance. (TARE)
0.0 g
CPA2P-F model:
Either position the standard pan
(8 g) or the filter pan (5 g) along
with the cover (3 g) beforehand.
2. Start calibration. (CAL)
C
The internal weight is CAL
applied automatically.
3. The balance is calibrated
0.2 g
(displayed only if menu CAL ± G
code 1 10 2 is set).
4. If the “Calibrate, then auto
Adjust*
adjust" setting is selected CAL
(1 10 1), the balance is
now adjusted automatically.
5. The calibration sequence
CC
is completed. CAL
6. The internal weight is removed.
0.0 g
* = Adjust displayed only if menu item 1 10 2 is selected.