Example: Counting parts of equal weight
Settings:
Menu: Counting program (menu code 2 1 4)
Step Key (or instruction) Display/Data output
1. Place empty container
+ 22.6 g
on the balance/scale.
2. Tare the balance. (TARE)
0.0 g
3. Add reference sample quantity
to container
(in this example: 10 pcs).
4. Initialize the balance. (F)
ref 10 (briefly)
+ 2.14 g
+ 10 pcs
nRef + 10 pcs
wRef + 2.14 g
5. Add uncounted parts + 500 pcs
as desired.
6. Print piece count, if desired. (A)
Qnt + 500 pcs
7. Display weight. (F) + 1070.0 g
8. Display piece count. (F)
+ 500 pcs
9. Unload the balance.
0 pcs
10. Repeat as necessary,
starting from Step 5.
11. Delete reference
sample quantity (CF)
0.0 g