Internal Calibration/Adjustment
(Only on Models with a Built-in Motorized Calibration Weight)
Models with a resolution of 0.1 mg, ED...-CW, GK.., GW... models, and verified models
(with the ...CE suffix) are equipped with a built-in motorized calibration weight as
a standard feature
Set the following parameters:
Setup: Bal.Scal. : Cal.-Adj.: Cal.Int. (menu code 1.1. 9. 4)
The built-in motorized calibration weight is applied and removed automatically for inter-
nal calibration.
Step Key (or instruction) Display
1. Tare the balance/scale
w 0.0 g
2. Start calibration
Q Cal.Int.
The built-in weight is
Cal. run.
applied automatically
3. Calibration/adjustment executed
Cal.end
4. The built-in weight is removed 0.0 g