EasyManua.ls Logo

Sartorius ME215S - Output Format (Line Format)

Sartorius ME215S
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
135
ID code characters I
1
)
Stat Status
ID Identification (identifier)
L ID Weighing series no. (lot)
W ID Weight set number
Nom Exact calibration weight
(value)
S ID Sample ID
NUM Numeric input
T1
Application tare memory 1
N Net weight (T1 = 0)
N1 Net weight (T1 0)
Qnt Piece count
Prc Percent
nRef Reference sample quantity
pRef Reference percentage
wRef Reference piece weight
Wxx% Reference percentage weight
mDef Target value for animal
weighing
Mul Calculated result
in animal weighing
x-Net Result in animal weighing
x-Res Calculated result in animal
weighing
Res Result using equation
(calculation)
Setp Target value for
checkweighing
Min Lower limit for checkweighing
Max Upper limit for checkweighing
Time Time that a value was stored
Compxx Component no. xx in formula-
tion
Tot.cp Total weight in formulation
n Transaction counter
Total Sum of all values
Avg. Average in statistics
s Standard deviation
srel Variation coefficient
Diff Difference between maximum
and minimum
Example: output weight of + 111.25507 mg
Position 1 2345678910111213141516
+111. 25507*mg*CRLF
Position 1: Plus + or minus sign – or space
Position 2: Space or weight value digit
Positions 3–10: Weight with decimal point; leading zeros are output as spaces
Position 11: Space
Positions 12–14: Characters for unit of measure or space
Position 15: Carriage return
Position 16: Line feed
Data Output Format with 22 Characters
When data with an ID code is output, the ID code consisting of 6 characters precedes the
data with the 16-character format. These 6 characters identify the subsequent value.
12345678910111213141516171819202122
IIIIII+DDDDDDDDD * UUUCRLF
*****–......... ***
**********
I: ID code character
1
) U: Unit symbol
1
) see “Toggle
*: Space between Weight Units”
D: Digit or letter CR: Carriage return
LF: Line feed
1
) depends on balance type; e.g., not all units and characters are
available on balances verified for use in legal metrology
Special Codes
12345678910111213141516171819202122
Stat****** * *––***** *CRLF
HH
LL
C
*: Space L: Underload
–: Final readout mode
(unstable weight) L L: Underload in checkweighing
H: Overload C: Calibration/adjustment
H H: Overload in checkweighing Draft shield and ionizer status; similar
to data output format with 16 characters
Error Codes
12345678910111213141516171819202122
Stat*****Er r *###****CRLF
*: Space # # #: Error code number

Table of Contents