Preparation
Select the balance function for “calibration/adjustment”: press s
To select the
Balance/scale functions: press the O soft key
Select
Calibration/adjustment: press the O soft key
Calibration/ CAL key function Ext. cal./adj.: default wt.
adjustment Ext. cal./adj.: user-defined wt.
Internal cal./adjustment
Internal linearization
Key blocked
reproTEST
ο Selection mode
Cal/adjustment ο Calibrate, then auto adjust
sequence Calibrate, then manual adjust
isoCAL function Off
Only adjustment prompt
On and reset application
isoCAL and linearization on*
ο On without resetting app.
Start automatic User-def. Adj. time1
adjustment adjustment times Adj. time2
Adj. time3
ο isoCAL
Print GLP/GMP ο Automatic if GLP is selected
adjustment record On request, from record memory
Parameter for Weight ID (Wt. ID)
external weight Cal./adj. wt.:
ο = factory setting
* this deletes (resets) the application
Save settings and exit Setup menu: press the
oo soft key
Releasing Access to
External Calibration
Remove the cover plate from the back
of the balance housing
Move switch 1 in the direction
of the arrow
> Switch left:
external calibration accessible
Switch right:
external calibration blocked
> Note:
Do not move switch 2