EasyManuals Logo

Sartorius Midrics 1 User Manual

Sartorius Midrics 1
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
Example:
External calibration and manual adjustment with default weights
(weighing parameters: factory settings)
(
1. Zero the scale.
J
2. Start calibration
(e.g., after calibration prompt:
flashing
WP symbol).
c.ext.def is shown for
two seconds.
You are prompted to place
the required weight on the
platform (e.g., 10,000 kg)
3. Position the calibration weight
on the weighing platform
The difference between
measured value and the true
mass is shown with a plus
or minus sign.
Ext. calibration Printout is generated,
Nom. + 10000 g if adjustment was not
Diff. + 1 g performed and the procedure
was cancelled by pressing (.
)
4. Start adjustment
(or cancel calibration/adjust-
ment by pressing ().
After adjustment,
the calibration weight value
is displayed.
-------------------- A GMP-compliant printout
24.10.2006 10:15 is generated.
Typ MW1P1
Ser.no. 12345678
Vers. 1.0103.11.2
BVers. 01-26-06
--------------------
Ext. calibration
Nom. + 10000 g
Diff. + 1 g
Ext. adjustment
Diff. + 0 g
--------------------
24.10.2006 10:15
Name:
--------------------
54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Midrics 1 and is the answer not in the manual?

Sartorius Midrics 1 Specifications

General IconGeneral
BrandSartorius
ModelMidrics 1
CategoryAccessories
LanguageEnglish

Related product manuals