Configuring the A/D Converter
Open the menu access switch
§ Remove the cap that covers the menu access switch on the left-hand side of the back
of the indicator
§ To do this, move the switch to the left (towards the interface connectors).
(“open position”)
Activate the service mode (see page 13)
)
Select weighing platform and confirm: Press the ) key
)
To select the menu item ADC configuration, press the k key several times until ADC-CON
appears. To confirm the menu item ADC configuration, press ).
k repeatedly
)
Select whether a standard configuration (STAND) or a verifiable configuration (VERIF.)
should be carried out (in this example, standard configuration).
See the next page for a detailed description of the procedure
Once you have completed the configuration, save the data using the menu item SAVE.
The A/D converter can now be treated like any standard weighing platform
in connection with the load sensor.
Close the menu access switch
Once ADC configuration has been completed, an adjustment of the weighing platform
(calibration/adjustment and linearization) must be carried out (see page 21
“Calibration/adjustment” and page 23 “Calibration without weights”)
wp-1
ADC-CON
STAND.
SAVE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Midrics 1 and is the answer not in the manual?

Sartorius Midrics 1 Specifications

General IconGeneral
BrandSartorius
ModelMidrics 1
CategoryAccessories
LanguageEnglish

Related product manuals