Error messages are shown in the main display. Error messages are shown permanently; Information messages are shown for 2 seconds.
After this the program automatically returns to the weighing mode.
Display Cause Remedy
ERR101 Key is stuck Release key or
Key pressed whilst switching on Contact your customer service partner
ERR 320 Operating program memory faulty Contact your customer service partner
ERR 340 Incorrect operating parameter (EEPROM) Switch scale off and then on again,
if Err 340 is permanently displayed:
Contact your customer service partner
ERR 341 Loss of data Contact your customer service partner
ERR 343 Loss of data in storage for Contact your customer service partner
external Alibi memory transaction number
Inf 01 Data output not compatible with Set output format correctly
output format
Inf 02 Calibration condition was not met Do not carry out calibration until after 0 display
e.g. not tared or Unload the weighing equipment
weighing pan loaded Tare with the ) key
Inf 03 Calibration could not be completed Keep to warm-up time and
within a certain time calibrate again
Inf 07 Last operating function is not permissible For setting changes please
use in legal metrology Contact your customer service partner
Inf 08 Load on the scale is too heavy for scale to be Please check if point 1.12 “Zero-setting
zeroed. range” was adhered to in your configuration.
Inf 09 Taring not possible for gross zero Zero the scale
Inf 10 Taring not possible with occupied tare memory Taring only possible after deletion of the
application program
Inf 22 Reference takeover error, Place a heavier weight on the platform
weight too small
Inf 23 Error in initializing an application Contact your customer service partner
Inf 29 Minimum load not achieved Reduce minimum load
(under Application, menu item 3.6)
Inf 71 Saving of measured value (or input) not possible None
(e.g. control limit too small or too large)
Inf 72 Saving of measured value not possible None
(e.g. maximum for item counter reached)
Inf 73 Stored data is deleted or cannot be read Contact your customer service partner
Inf 74 Function is locked None
(e.g. menu is locked)
Inf 98 No weighing platform connected Contact your customer service partner
Inf 99 No weighing platform connected Contact your customer service partner
NO WP No weighing platform connected Connecting weighing platform