8
You can calibrate/adjust the scale by pressing the tare
key U (4).
After connection to AC power and before each
calibration/adjustment, allow the scale to warm up for
approximately 30 minutes.
§ Hold down the tare key U (4) for 2 sec.
When 5000 is displayed, release the key.
You can select the calibration weight via the I
toggle key (6).
Example: calibration weight 5000 g
§ Confirm the calibration weight selected by pressing
the L key [ENTER] (5)
§ After the scale has been tared, place the displayed
calibration weight on the weighing pan.
§ Center the calibration weight on the weighing pan (10).
Calibration/adjustment is performed automatically.
When the calibration has been accepted, the symbol
will disappear and the stability symbol is displayed
§ After calibration and adjustment, remove the weight.
Calibration/Adjustment