EasyManua.ls Logo

Sartorius PMA.Quality PMA7501H - Calibration;Adjustment

Sartorius PMA.Quality PMA7501H
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration/Adjustment
12
You can calibrate/adjust the scale by pressing the
U key (5).
To do so, menu code 1 5 1 must be selected.
Calibration weight: 5,000 g; accuracy: + 0.075 g.
After connection to AC power and before each
calibration/adjustment, allow the scale to warm up
for approx. 30 min.
Hold down the U key (5) for 2 sec. When 5000 is
displayed, release the key.
Center the calibration weight on the weighing pan
(11).
Calibration/adjustment is performed automatically.
After calibration and adjustment, remove the weight.

Related product manuals