82 Installation and Operating Instructions Secura, Quintix, Practum
Technical Specifications
Verified Models with Country-specific Type Approval Certificate
Secura
®
Model 224-
1x
2
)
124-
1x
2
)
1103-
1x
2
)
613-
1x
2
)
513-
1x
2
)
313-
1x
2
)
213-
1x
2
)
6102-
1x
2
)
5102-
1x
2
)
3102-
1x
2
)
Accuracy class
kkk K K KK K K K
Type
3
) SQP-A SQP-A SQP-I SQP-B SQP-B SQP-B SQP-B SQP-C SQP-C SQP-C
Max g 220 120 1,100 610 510 310 210 6,100 5,100 3,100
Scale interval d mg 0.1 0.1 1 1 1 1 1 10 10 10
Verification scale interval e mg 1 1 10 10 10 10 10 100 100 100
Min mg 10 10 100 20 20 20 20 500 500 500
Min (only for Models ...-1OIN) mg 100 100 – 200 200 200 200 5,000 5,000 5,000
Tare (subtractive)
< 100% of max. weighing capacity
Minimum sample weight
as per USP, typical
g 0.12 0.12 1.5 1.5 1.5 1.5 1.5 12 12 12
Typical stabilization time s 2 2 1.5 1 1 1 1 1 1 1
isoCAL:
– Temperature change
– Time interval
K
h
1.5
4
1.5
4
1.5
4
2
6
2
6
2
6
2
6
2
6
2
6
2
6
Display result
(depending on the set filter level)
s 0.2 0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm d 90 d 90 d 120 d 120 d 120 d 120 d 120 d 180 d 180 d 180
Weighing chamber height* mm 209 209 209 209 209 209 209 – – –
Net weight, approx. kg 5.1 5.1 5.9 5.1 5.1 5.1 5.1 5.2 5.2 5.2
Model 2102-1x
2
) 1102-1x
2
) 612-1x
2
) 6101-1x
2
) 3101-1x
2
)
Accuracy class
KKK KK
Type
3
) SQP-C SQP-C SQP-C SQP-C SQP-C
Max g 2,100 1,100 610 6,100 3,100
Scale interval d mg 10 10 10 100 100
Verification scale interval e mg 100 100 100 100 100
Min mg 500 500 500 5,000 5,000
Min (only for Models ...-1OIN) g 55555
Tare (subtractive)
< 100% of max. weighing capacity
Minimum sample weight
as per USP, typical
g1212126060
Typical stabilization time s 11111
isoCAL:
– Temperature change
– Time interval
K
h
2
6
2
6
2
6
2
6
2
6
Display result
(depending on the set filter level)
s 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm d 180 d 180 d 180 d 180 d 180
Net weight, approx. kg 5.2 5.2 5.2 5.2 5.2
Model-specific Data