7. Troubleshootin
Fault Possible Cause Solution
Tip not tightened correctly. Re-tighten the tip.
Defective handpiece. Replace the handpiece.
Oxidization and/or black marks on the pins of the
connector.
Clean the contacts or replace the cord or return to Satelec After Sales Service.
No ultrasound
Cord cut. Replace the cord or return to Satelec After Sales Service.
Water leaking from the
handpiece connector.
Worn handpiece o'ring. Replace the o'ring.
Leaking from the end of the
tip, pedal released.
Dirt in the water circuit (electric valve blocked open)
Clean the circuit and electric valve (check the electric valve is working properly)
or return to Satelec After Sales Service.
Blown bulb. Replace the bulb.
Incorrect use or setting of the mode. Refer to the instructions.
No light with the Lux
handpiece.
Electrical problem. Return to Satelec After Sales Service.
No air in the supply. Contact your approved installer.
Overheating of the Lux
handpiece.
Air inlet pipe not connected. Connect the air inlet.
No air or water coming from the pipes connected to the
mains supply.
Contact your approved installer.
No powder, air or water.
Handpiece and / or nozzle blocked.
Disconnect the handpiece and check that air and powder are coming out of the
cord:
- If not, return the device to Satelec After Sales Service.
- If they are, unscrew the handpiece nozzle, unblock the nozzle with a file and
place the handpiece in an ultrasound tank for 10 minutes.
Powder: drum empty or too full. Fill the drum or remove the excess.
No powder or water but air
present.
Water: water inlet not connected or incorrect water
setting.
Check the water inlet pipe or adjust the water flow on the handpiece cord.
Incorrect use or setting of the mode. Refer to the instructions.
Exiting from energy saving mode not possible. Return to Satelec After Sales Service.
Energy saving mode not
working properly.
Adjustment impossible. Return to Satelec After Sales Service.