3.4. Adjusting the pump flow
1
2
The pump flow is pre-adjusted in the factory and can be
modified to suit user requirements.
Using a screwdriver turn the screw (1) anti-clockwise
by 360 °.
Using flat pliers turn the marking of the slide (2) to the
desired position.
Tighten the screw.
Examples:
external 80
%
internal100
%
internal 20 %
A
B
D
C
Internal applications in the bath
A 100 % internal bath circulation
(for large bath tanks)
B Reduced internal bath circulation
(for smooth surface of bath fluid)
External/internal applications
C 40 % external discharge,
60 % internal circulation
(for large bath tanks)
D 80 % external discharge,
20 % internal circulation
(for small bath tanks)
Caution: Securely attach all tubing to prevent slipping.
4. Operating procedures
4.1. Power connection
Caution:
Only connect the unit to a power socket with earthing contact
(PE – protective earth)!
We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the supply
voltage specified on the identification plate.