SHC 2000
9.900.900.719 SHC2000-manual-111609 21
7. Safety recommendations
Follow the safety recommendations to prevent damage to persons
or property. Further, the valid safety instructions for working places
must be followed.
• Only connect the unit to a power socket with earthing contact (PE –
protective earth)!
• Place the instrument on an even surface on a pad made of non-
inflammable material.
• Do not stay in the area below the unit.
• Make sure you read and understand all instructions and safety
precautions listed in this manual before installing or operating your
unit.
• Observe the fire point of the bath medium used.
The excess temperature protection should be set at least 25 °C below
the fire point.
• Never operate the unit without bath fluid in the bath.
• Pay attention to the thermal expansion of bath oil during heating to
avoid overflowing of the fluid.
• Prevent water from penetrating into the hot bath oil.
• Exercise caution when emptying hot bath fluids!
Check the temperature of the bath fluid prior to draining.
• Observe the limited working temperature range when using
plastic bath tanks.
• Employ suitable connecting tubing.
Make sure that the tubes are securely attached.
• Never operate damaged or leaking equipment.
• Always turn off the unit and disconnect the mains cable from the
power source before performing any service or maintenance
procedures, or before moving the unit.
• Always empty the bath before moving the unit.
• Never operate equipment with damaged mains power cables.
• Some parts of the bath cover and the pump connections may
become extremely warm during continuous operation. Therefore,
exercise particular caution when touching these parts.