SHC 2000
9.900.900.719 SHC2000-manual-111609 19
Notice:
Check the safety installations at least twice a year!
Excess temperature protection according to IEC 61010-2-010
With a screwdriver turn back the adjustable excess temperature
protection until the shut-down point (actual temperature).
Low level protection according to IEC 61010-2-010
To check the function of the float, it can be manually lowered with a
screwdriver for example.
6. Troubleshooting guide / Error messages
+8
Whenever the microprocessor electronics registers a failure, a
complete shutdown of the heater and circulating pump is
performed. The alarm light "
" illuminates and a continuous
signal tone sounds.
The LED temperature display indicates the cause for the alarm in
form of a code.
Press enter
to quit the audible signal.
The circulator is operated without bath fluid, or the liquid level
is insufficient.
Replenish the bath tank with the bath fluid.
Tube breakage has occurred (insufficient filling level due to
excessive bath fluid pumped out). Replace the tubing and
replenish the bath tank with the bath fluid.
Cable of the working temperature sensor interrupted or short-
circuited.
Defect of the working or excess temperature sensor.
Working temperature and excess temperature sensors report
a temperature difference of more than 25 °C.
Error in A/D converter