SHC 2000
16
1. Turn on the circulator with the mains switch.
2. Start the unit
Press enter
for about 4 seconds.
3. Start the calibration:
Press the hidden key and the
key at the same time.
The controller attempts to heat up to the working
temperature (30 °C).The LED temperature DISPLAY
indicates the actual bath temperature.
4. When the working temperature is reached, wait until the
circulator maintains a constant bath temperature
(after about 3 minutes).
28.87
°C
5. Read the bath temperature from the calibrated
thermometer and set the rounded value on the circulator.
(example: 28.87 °C rounded to 28.9 °C).
Setting the temperature
Press one of the keys for a short moment.
The setpoint value instead of the actual value is
indicated on the display. The value can now be
changed.
Change value:
Press to set a higher value.
Press to set a lower value.
Keep the keys depressed for the value to change fast.
Press enter to store the value.
For a short while the LED temperature DISPLAY
indicates the message „CAL“.
L Notice:
If the determined value is bigger than the tolerance limit
of ±5 °C, the value is ignored for the input!