SHC 2000
8
Safety installations according to IEC 61010-2-010:
Excess temperature protection adjustable from 0 °C ... 170 °C
Low liquid level protection float switch
Classification according to DIN 12876-1 class III
Alarm message optical + audible (permanent)
Environmental conditions according to IEC 61 010-1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C (for storage and transportation)
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Protection class according to IEC 60 529 IP21
Power supply: corresponds to Class I; according to VDE 0106 T1
not for use in explosive atmosphere
Max. mains fluctuations of ±10 % are permissible.
Overvoltage category II
Pollution degree 2
Standards for interference resistance
EN 61326: 1997 + A1: 1998 + A2: 2001 + A3: 2003
Emitted interferences
The unit adheres to the threshold values for emitted interferences
according to table 3.
Interference resistance
The unit conforms to the requirements according to table B.1.