humimeter RH2 aW value measuring chamber user manual
Page 34
Typical sorption isotherms of various foods and materials taken from literature:
7.3 Application range
Within the normal application range (normal range) the accuracy of the device is as
indicated. A long-term application beyond the normal application range (max. range),
particularly at an air humidity of more than 80%, can lead to higher measuring errors
(+3 % after 60 hours). Back in the normal application range, the sensor will return to
the indicated accuracy automatically.