Temperature Humidity standards
35 % 50 % 80 %
20 °C 34.8 % 49.8 % 79.8 %
21 °C 34.9 % 49.9 % 79.9 %
22 °C 34.9 % 50.0 % 79.9 %
23 °C 35.0 % 50.0 % 79.9 %
24 °C 35.1 % 50.1 % 80.0 %
25 °C 35.2 % 50.2 % 80.0 %
26 °C 35.3 % 50.2 % 80.0 %
6. Note down the real humidity value.
7. Compare the noted displayed measuring value with the real humidity value.
» If the indicated value shows a difference higher than 1.5 % relative humidity, it is
recommended to perform a recalibration. A calibration can be performed by the
company Schaller Messtechnik GmbH.
» It is also possible to run the calibration by yourself using the calibration am-
poules: For instructions on how to perform calibration, please contact Schaller
Messtechnik GmbH or your distributor.
8. Open the measuring chamber and clean the glass jar with plenty of water. The
textile pad must be washed out well and can then be dispossed.
13. Faults
If the measures listed below fail to remedy any faults or if the device has faults not
listed here, please contact Schaller Messtechnik GmbH.
Fault Cause Remedy
Measuring error The temperature is outside the
operating temperature: lower
than -10 °C or higher than
+60 °C
Only use the device in tem-
peratures between -10 °C
and +60 °C.
Measurement error due to too
short temperature adjustment
time
Let the device adjust to
the surroundings (see "5.3
Adjustment behaviour of
the sensor").