humimeter RH2 aW value measuring chamber user manual
Page 50
Fault Cause Remedy
Wrong product type Check whether you have
selected the right product
type before taking a reading
(see "7. Product types").
Filling volume too low Fill the screw-top jar to at
least two thirds with fine
material or at least half-way
with coarse material.
Dripping water or sprayed
water
Direct contact of the sensor
with dripping or sprayed
water will destroy it.
Irreversible damage of the sen-
sor due to aggressive gases
Please contact your dealer.
Condensation caused by a
change in temperature
Condensation on the sensor
interferes with the calibra-
tion. Let the device adjust
to the surrounding tempe-
rature.
Polluted sensor Please contact your dealer.
Foreign particles on the sensor Please contact your dealer.
Data transfer to Log-
Memorizer failed
Interface has not been con-
figured
The interface only has to
be configured once. To do
so, press the F1 key on your
computer and read the Help
file for your LogMemorizer
program.