Operation
107
Note
If error message

appears see chapter 6 W
HAT TO DO IF...
AutoRead During calibrating, the
AutoRead
function is automatically
activated. The AR display flashes. The calibration procedure
is finished when the AR display stops flashing.
4 Immerse the measuring cell into the 0.01 mol/l KCL
control standard.
5 Press the <run/enter> key.
If no temperature probe is connected, enter the
current temperature of the solution using <
> <>
and confirm with <run/enter>
If a temperature probe is connected, the AR
measurement to determine the cell constant starts.
flashes until a stable signal is achieved. The
determined cell constant ist displayed; the
measuring instrument automatically stores the cell
constant.