SCIEX ExionLC - Error Messages; Communication, Calibration, and Cell Errors; Lamp, Temperature, Sensor, and System Errors

SCIEX ExionLC
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The following table shows the error messages that can appear on the acquisition computer, along with their
probable causes and corrective action.
Corrective ActionProbable CauseError MessageError Code
Verify the operating state of the
acquisition computer and the
LAN cable connection.
The ethernet communication with
the computer timed out.
TCP/IP communications timed
out.
0x8952
Verify the operating state of the
controller and the LAN cable
connection.
The ethernet communication with
the controller timed out.
CBM LAN timeout.0x8954
Perform cell registration,
exposure time adjustment, and
wavelength calibration.
An unregistered cell was installed
in the module.
Cell is not calibrated.0x8965
Verify the connection of the cell
cable.
The cell ID could not be read.
Set the exposure time.The exposure time of the installed
cell is not adjusted.
The exposure time of the not
adjusted. Set the exposure
time.
0x8966
Perform wavelength calibration.The wavelength of the installed
cell is not calibrated.
The wavelength of the
installed cell is not calibrated.
Perform wavelength
calibration.
0x8967
If the filter is clogged, replace
the filter.
The dust filter on the back of the
module is clogged.
Temperature is increasing
inside the module.
0x89D2
Inspect the tubing. Refer to
Corrective Action against
Leaks from Cell Window
and Clean the Flow Cell.
The mobile phase is leaking inside
the system.
Leak is detected.0x8921
Replace the D2 lamp. Refer to
Replace the D2 Lamp.
The D2 has lamp malfunctioned.D2 lamp is over current.0x89C2
Turn off the power and then
contact the FSE.
The lighting circuit has
malfunctioned.
ExionLC™ PDA DetectorOperator Guide
41 of 47RUO-IDV-05-1842-B
A
Error Messages