436-GC/456-GC
510 Error codes
3
If the used gas is Hydrogen (H
2
) the Error level is Error, in all other cases an Alert is given.
Page: 66
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
Code Error Description Error level Cause Condition(s) GC action Action needed
more then 10 sec.
(gas-supply)
306 Front DFC Chan3 out of range Alert/Error
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Activate Safety Method
Turn off/on flow, check source
(gas-supply)
307 Middle DFC Chan1 out of range Alert/Error
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Activate Safety Method
Turn off/on flow, check source
(gas-supply)
308 Middle DFC Chan2 out of range Alert/Error
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Activate Safety Method
Turn off/on flow, check source
(gas-supply)
309 Middle DFC Chan3 out of range Alert/Error
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Activate Safety Method
Turn off/on flow, check source
(gas-supply)
310 Rear DFC Chan1 out of range Alert/Error
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Activate Safety Method
Turn off/on flow, check source
(gas-supply)
311 Rear DFC Chan2 out of range Alert/Error
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Activate Safety Method
Turn off/on flow, check source
(gas-supply)
312 Rear DFC Chan3 out of range Alert/Error
Flow is above max +10 % or below -5 mL/min for
more then 10 sec.
Activate Safety Method
Turn off/on flow, check source
(gas-supply)
313
Front EFC Channel 1, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
314
Front EFC Channel 2, flow ready time out
Alert
Error
3
Unit not ready for 10 minutes
-
-
315
Front EFC Channel 3, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
316
Middle EFC Channel 1, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
317
Middle EFC Channel 2, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
318
Middle EFC Channel 3, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
319
Rear EFC Channel 1, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
320
Rear EFC Channel 2, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
321
Rear EFC Channel 3, flow ready time out
Alert
Error
Unit not ready for 10 minutes
-
-
NPD Bead
332
Front NPD bead Shorted
333
Middle NPD bead Shorted
334
Rear NPD bead Shorted
335
Front NPD bead Open
336
Middle NPD bead Open
337
Rear NPD bead Open
339
DataStream Manager Stopped Incorrect
Local Automation
340
Local/Automation Injection Position Incorrect
341
Local/Automation First Vial Incorrect
342
Local/Automation Last Vial Incorrect
343
Local/Automation First Injection Volume Incorrect
344
Local/Automation Second Injection Volume Incorrect
368
Password Incorrect
Auto Sampler (AS)
370
Sampler old firmware
Event
Syringe stepper motor firmware old
-
-
371
Sampler Faulted
Alert
Mismatch AutoSampler
Notification Message
Cable in SID2, check power