KeyPad service manual
Rev.1.1
21
o Press the LEFT ARROW key a couple of time and see if the cursor
switched from ON to OFF and so on.
o Press the RIGHT ARROW key a couple of time and see if the
cursor switched from ON to OFF and so on.
Select HEATER OFF and press ENTER (Make sure that a DigiPREP is
connected
Enter a time scale of 1 minute and press enter
Press the ALARM, TARGET TEMPERATURE and TARGET TIME button one
at a time and verify if they are correctly doing what they are supposed to do:
o ALARM: A small “AL” should appear on the LCD display.
o TARGET TEMPERATURE: The target temperature menu should
appear.
o TARGET TIME: The target time menu should appear.
3.2 KeyPad Diagnostic Section
3.2.1 Entire row or column problem
If all buttons in a row or column are failing, with the helps of figure 9, identify
the corresponding pull-up resistor.
o Remove the CPU board from the main PCB
o Measure with a multimeter the value of the corresponding resistor.
o If it is 4.7K, try to verify the connection (soldering) between the
VCC and the resistor.
o Then check if the resistor is also connected to each button
corresponding to its row or column (see figure 10 for resistors
location)

Table of Contents

Other manuals for SCP SCIENCE DigiPREP HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SCP SCIENCE DigiPREP HP and is the answer not in the manual?

SCP SCIENCE DigiPREP HP Specifications

General IconGeneral
BrandSCP SCIENCE
ModelDigiPREP HP
CategoryKeypad
LanguageEnglish

Related product manuals