EasyManuals Logo

SCP SCIENCE DigiPREP HP Service Manual

SCP SCIENCE DigiPREP HP
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
KeyPad service manual
Rev.1.1
31
o Measure with a multimeter the voltage between T2 test point
and GND Jumper (analog ground).
o If there is a pulse or a steady 5V then the Q3 switch is fine.
The problem is probably the solid-state relay.
o If there is always 0V and the LED is ON or toggles, then
replace Q3.
Solid State Relay verification
o The only way to see if the SSR is behaving accordingly is by
measuring between QD6 and QD5 (see figure 19 for
location). Make sure that there is 120V/230V AC (Toggling
or not). Also refer to figure 19 for an example of what can
be seen on a scope if the control is toggling.
o Make sure that the heating block temperature is lower than
the target temperature if you want the SSR to toggle.
NOTE: If you’re using a scope, it must be isolated from the wall
outlet.
Figure 19: SSR test pins location
Legend:
1. QD5
2. QD6
1
2