1
DigiPREP Digestion Blocks
Version 16.0
INTRODUCTION
DigiPREP Digestion blocks are designed for labora-
tory use in the digestion and evaporation of a variety of
samples, primarily water, wastewater, soil, and sludge
samples. They are Teflon
®
coated.
DigiPREP Jr, DigiPREP MS, DigiPREP LS and
DigiPREP HP blocks have a temperature range of
ambient to 180 °C. These blocks are designed to
hold DigiTUBE sample digestion vessels and glass
/ quartz digestion tubes.
DigiPREP 240 Jr, DigiPREP 240 MS and
DigiPREP 240 LS blocks have a temperature range
of ambient to 240 °C. A “spill proof” design prevents
acid migration to the inside of the block housing. The
top white ring is removable.
Optional accessories include DigiPROBE, which is
used to monitor sample temperature, and DigiSET,
which is used to monitor and shut off the heating
system at a set sample volume level.
Old controllers will work with new
blocks but will not allow the blocks
to reach 240 °C. New controllers are
required to reach 240 °C.
DigiTUBEs have a temperature
range of up to 115 °C.
UT