OPERATIONAL PRECAUTIONS
DigiPREP block must be connected to a
controller that is properly grounded. In the event
of an electrical short circuit, grounding reduces
the risk of electrical shock by providing an escape
wire for electrical current. DigiPREP controllers are
supplied with an electrical cord that has a grounding
wire with a grounding plug. Ensure that the plug
is connected into an outlet that has been properly
installed and grounded. Consult a qualified
electrician if there is any doubt about the existence
of proper grounding at the outlet.
DigiPREP block is designed to be lifted from
the bottom only. Ensure that the controller is
disconnected prior to moving the block. Do not
attempt to lift the block by the top white trim
as damage may result.
DigiPREP blocks are elevated-temperature
digestion systems with exposed hot surfaces. Be
careful around the instrument when it is in operation.
SAFETY INFORMATION
A minimum of safety considerations must be
followed when working with DigiPREP low
temperature blocks in order to maintain good
laboratory practice:
` Do not place the DigiPREP blocks on a sur-
face containing flammable material.
` DigiPREP block requires a clearance of three
inches (7.5 cm) on all sides.
` DigiPREP block should be operated in a fume hood.
` Do not place the DigiPREP controller in a fume
hood.
` DigiPREP system must be plugged into a
three-prong grounded outlet.
For indoor use only. Disconnect
mains power supply before any
services.
Mains power source suitable for
equipment shall be provided at the
installation location.
If the equipment is used in a manner
not specified by
SCP SCIENCE,
the protection provided by the
equipment may be impaired.
INSTALLATION
UNPACKING
Retain all packing material in case
you need to return the instrument to
the manufacturer for service.
Verify that all components listed have been
included:
` DigiPREP block
` Operation manual
` DigiPROBE dongle
(ambient to 180 °C series only)
` DigiTUBE rack:
■ Jr / 240 Jr: 1 rack
■ MS / 240 MS: 2 racks
■ LS / 240 LS: 3 racks
■ HP: No racks
Optional items, such as glass,
quartz, calibrated or non-calibrated
digestion tubes may have been
included in the shipment. Carefully
check the packing list(s) and
the contents of shipping cartons
to verify that all items listed are
included. Please notify
SCP SCIENCE
or your local distributor of any
discrepancies.
Inspect the products for any shipping damage.
If the instrument has been damaged in shipping,
contact the freight carrier to report damage and
to file a damage report. Also contact the SCP
SCIENCE service department to report damage
and to request service information.
If you find evidence of damage, do
not attempt to operate the
instrument.