EasyManua.ls Logo

SCP SCIENCE NOVAWAVE - IR Calibration - Reading Data; IR Calibration - Sending Data to MicroB; IR Calibration - Saving Data; IR Calibration - Prerequisites

SCP SCIENCE NOVAWAVE
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Version 0.15.1
NOVAWAVE - A New Category of Automated Microwave Digestion | 31
Read New Data once the thermopile
calibrating device is connected, press this
button to calibrate the
NOVAWAVE
.
NOVAWAVE
is factory calibrated, and
should be calibrated every 6 months during
a preventive maintenance visit by a trained
Service Engineer.
Send to MicroB once the data is verified,
the information needs to be uploaded to the
CPU.
Save is used to save the data to the graph-
ical user interface (GUI).
Save to USB a log of the calibration data can
be stored to a USB key.
The values select displayed in P1 to P5 are
the target values of each IR sensor to. The
calibration is a 5 point calibration. The values
shown in the table are not representative of
a true calibration as the table. Values should
be
random.
IR Calibration
By pressing Read New Data the fol-
lowing prompt comes up.