EasyManuals Logo

SCP SCIENCE NOVAWAVE User Manual

SCP SCIENCE NOVAWAVE
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Version 0.15.1
44 |
9
Troubleshooting
No. Error Message Description
1 COM.ERROR Unrecognized packet.
2 ERROR! ID=Stalled DESC= Section 1 (2,3,4) is stalled. FA Rack is stalled at B-F section.
3 ERROR!!! ID= Timed Out. DESC= Section Operation has timed out. FA Operation B-F timed out. (R-L, F-B or L-R).
4 ERROR COM Timed out! FA No response from MB.
5 ERROR No OK received from MB. FA OK not received from MB.
6 Process terminated with exit code #. Press OK to restart. GUI crashed code number #. Press OK to resume.
7 DIGESTION.RUN TUBE 01 E_CODE=02 Never reached target temperature.
8 DIGESTION.RUN TUBE 01 E_CODE=03 Temperature rising too slowly in tube number #.
9 DIGESTION.RUN TUBE 01 E_CODE=04 Tube number # overheats.
10 MB_SYNTH System initialize.
11 ERROR Digestion Operation has timed out! Digestion operation has timed out on tube number #.
12 ERROR DIGESTION_ALARM_ERROR Magnetron is alarmed. Magnetron number # has an alarm.
13 DIGESTION_OVERHEAT_ERROR Magnetron overheated Magnetron number # overheats.
14 ACTUATOR_STALLED_ERROR Actuator is stalled.
15 ACTUATOR_OPERATION_TIMEDOUT_ERROR Actuator operation has timed out.
16 CAVITY_STALLED_ERROR Tunnel motor is stalled.
17 CAVITY_UNKNOWN_RACK Please remove unknown rack from tunnel.
18 CAVITY_OPERATION_TIMEDOUT_ERROR Tunnel operation has timed out.
19 CAVITY_OBSTRUCTED_ERROR Tunnel is obstructed.
20 CAVITY_NOT_INITAILIZED_ERROR Tunnel motors need initialization.
21 CAVITY_NOT_NOW_ERROR This command cannot be performed.
22 CAVITY_ORITENTATION_FAILED_ERROR Unknown rack orientation.
Messages may occur during system operation due to operator error or component failure. For further assistance, contact SCP SCIENCE
Service Department or your local distributor.
List of Error Messages:
23 CAVITY_INTERNAL_ERROR An error occurred in tunnel.
24 RACK ID FAILED
25 RACK IN REVERSE POSITION
26 PMB Thermal switch has failed.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SCP SCIENCE NOVAWAVE and is the answer not in the manual?

SCP SCIENCE NOVAWAVE Specifications

General IconGeneral
BrandSCP SCIENCE
ModelNOVAWAVE
CategoryLaboratory Equipment
LanguageEnglish