Application Note 0081
Rev 1: 30.09.2016
Page 1 of 2
How to Pair the PV200 or PV210 with the Survey 200R
The following document is intended to be a troubleshooting guide to ensure you are setting up your Seaward instrument(s) in
the correct manner. Please check all aspects of this before contacting the technical support helpline.
Please register your Seaward product at http://www.seaward.co.uk/userfiles/register-product/select-product.php to
activate your warranty, receive support and other benefits.
Pairing the Instruments
1) Before attempting to pair please make sure the batteries in both instruments are good.
2) Make sure there are no other units operating nearby.
3)
Make sure the PV200 / PV210 and Survey 200R instruments are turned off.
4) On the Survey 200R, press and hold the “On/Off” keys, keep both keys pressed.
Note: The Survey 200R will now send the pairing signal (beeping approximately every second).
5)
As soon as the buttons are pressed, and held on the 200R, press and hold the “Riso” and “Mode” keys on
the PV200 / PV210 - keep all buttons pressed.