Model 336 17
… Er:012: appears in the display?
– The scale was switched on with too great a load on it. Remove load and restart.
8. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal
requirements. The year of initial calibration can be found behind the CE symbol on the
rating plate above the number of the appointed office, 0109 (Hessen Weights & Measures
Office). Recalibration is essential if one or more security marks are damaged or the
calibration counter does not match the number on the applicable calibration counter mark
(cf. page 15).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
9. Technical data
Dimensions
length:300 mm
width:640 mm
height:105 mm
Weight approx. 3.5 kg
Temperature range +10 °C to +40 °C
Maximum load 15 kg
Graduations 10 g
Calibration as per Directive
2009/23/EC Class III
Accuracy at first calibration
up to 5 kg: ± 5 g
5 kg up to 15 kg: ± 10 g
Medical product
as per Directive
93/42/EEC Class I
Tare range up to 10 kg / 22 lbs
Height of figures 20 mm
Power supply Battery/
(option mains unit)
Power consumption approx. 25 mA