9
What do I do if …
… no weight is displayed when there is
a weight on the scale?
Is the scale switched on?
Check the batteries
0.000 does not appear before weigh-
ing?
Switch the scale off and back on again,
making sure there is no load on the scale.
… one segment comes on permanently
or not at all?
This segment is faulty. Inform Servicing.
appears in the display?
Battery power is dropping. You will need to
replace the batteries within the next few
days.
batt appears in the display?
The batteries are flat. Insert new batteries.
St0p appears in the display?
The maximum load has been exceeded.
appears in the display?
The ambient temperature of the scale is too
high or too low. Place the scale in an ambi-
ent temperature between +10 ˚C and
+40 ˚C. Wait about 15 minutes until the
scale has acclimatised to the ambient tem-
perature and then weigh again.
Er:011: appears in the display?
The scale is overloaded or the load is une-
venly distributed. Take the load off the scale
or distribute weight more evenly and restart
scale.
Er:012: appears in the display?
The scale was switched on with too great a
load on it. Remove load and restart.
Technical data
Dimensions
Depth: 358 mm
Width: 620 mm
Height: 190 mm
Weight approx. 3.2 kg
Temperature range +10 °C to +40 °C
Maximum load 20 kg / 44 lbs
Graduations
0 up to 10 kg /
0 up to 22 lbs: 5 g / 0.2 ozs
10 up to 20 kg /
22 up to 44 lbs: 10 g / 0.5 ozs
Accuracy better than ±0,3% / ±10 g
Tare range up to10 kg / 22 lbs
Auto Hold function yes
Power supply battery/
mains unit (option)
Power consumption approx. 25 mA
Accessories
Plug-in mains unit
Use only seca original plug-in mains units
with 9 V or controlled 12 V output voltage.
Commercially-available mains units may
supply a higher voltage than stated and
thus lead to damage to the scale.
Plug-type mains adapter, Euro: 230V~ /
50Hz / 12V= / 130mA,
order no. 68-32-10-252
Tabletop mains power unit, can be
changed over, Euro: 115V~ / 230V~ /
50-60Hz / 12V= / 350mA,
order no. 68-32-10-243
Plug-type mains adapter, USA: 120V~ /
60Hz / 12V= / 150mA,
order no. 68-32-10-259
Head and foot rest
order no. 418-00-00-009
Disposal
Batteries
Do not dispose of used batteries in house-
hold waste. Dispose of batteries via local
collection points. When buying new sup-
plies, buy batteries low in harmful substanc-
es which contain no mercury (Hg),
cadmium (Cd) or lead (Pb).
Disposing of the scale
If the scale can no longer be used, the re-
sponsible waste disposal association will be
pleased to give you information on how to
dispose of it properly.