EasyManua.ls Logo

Seca 376 - 9. WHAT TO DO IF...?; Troubleshooting: Weight Display & Zero Point Errors

Seca 376
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56 •
NOTE:
For further operation of the wireless printer see
its user manual.
8. CLEANING
To clean the tray and housing of the scales use a
household detergent or standard disinfectant as
required. Take note of the instructions provided by the
manufacturer.
9. WHAT TO DO IF …?
Malfunction Cause/Remedy
… no weight is dis-
played during weighing?
The scales are not connected to the power supply
- Check the scales are switched on
- Check batteries are inserted
0.000 does not appear
before weighing?
A load was placed on the scales before being
switched on.
- Remove load on scales
- Switch scales off and back on again
… a segment lights up
continuously or not at
all?
There is a fault at that point.
- Notify maintenance service.
is displayed?
Battery voltage is falling off.
- Insert new batteries as soon as possible
bAtt is displayed?
Batteries are flat.
- Insert new batteries
St0p is displayed?
The maximum load has been exceeded.
- Reduce load
is displayed?
The ambient temperature of the scales is too high
or too low.
- Set up scales in an ambient temperature
between +10 °C and +40 °C
- Wait approx. 15 minutes until scales have
adjusted to ambient temperature

Other manuals for Seca 376

Related product manuals